Cause is most likely due to highly saturated ion front (a long stretch in the dye at low MW where it’s flat and around 32,700 RFU
-
- Open the raw data folder at the top of the file name tree
- Zoom into the area between the ion front and the first ILS peak
- Take note of the raw data frame number (X-axis) in this region
- Return to the analysis template, second screen (Data Process Options)
- Uncheck Auto-Range (frame) and enter the number noted in iv.
- Complete the analysis using this setting